• Very easy operation with 4-button control and integrated reference temperature sensor
• PC interface with connection cable to USB for use with SIKA calibration software
• Low weight and stable handle for easy transport
• Optional accessory: Transport case with or without trolley
• Qualified for SIKA Gold Service
• Especially suitable for calibration of devices under test with special sensor geometry. DUTs are simply immersed in the calibration liquid, no calibration insert with specific bore hole pattern is required
Made in Germany. brand SIKA, model TPM165S
International delivery to United Arab Emirates, Kuwait, Qatar, Saudi Arabia, Saudi Arabia, Bahrain, Oman and other Middle Eastern countries.
Our warehouses are located in UAE.
price by request
Available
Only the best I&C brands
Our company is a UAE supplier of control and measuring instruments from well-known world brands
Discounts for organisations
We sell high-tech equipment at best prices. We provide really advantageous conditions for partners
Fast delivery across the Middle East
Supply to United Arab Emirates, Kuwait, Qatar, Saudi Arabia, Bahrain, Oman and other countries
Professional advice on device selection
Our specialists are ready to advise you on any technical question. We can find an analogue of any device
Whether for on-site use in the workshop, in the measuring and test room or laboratory, SIKA temperature calibrators can be used everywhere. A wide range of industries and various application areas are covered.
• assembly, commissioning
• manufacturing, production
• maintenance, service
• quality assurance, test equipment monitoring
• repair
Calibration bath function
If temperature sensors with unusual shapes and dimensions are to be checked, the use of calibration liquids offers advantages. The DUT is placed directly into the liquid without insulating air gaps, thus creating a direct temperature seal between the calibrator and the DUT. The liquid, e.g. silicone oil, is selected according to the required calibration temperature. The continuously variable control of the magnetic stirrer, together with the pull-out sensor basket, causes vortices in the calibration liquid and thus creates a large homogenous temperature zone. Furthermore, the sensor basket ensures an unobstructed stirring operation and serves to protect the tank bottom.
TP M165S | ||
Temperature range | -35…155 °C at ambient temperature 20 °C | -31…311 °F at ambient temperature 68 °F |
Calibration volume | Ø 60 x 170 mm | |
Calibration bath | ||
Accuracy | ±0.1 °C | ±0.18 °F |
Temperature stability | ±0.05 °C | ±0.09 °F |
Resolution of the temperature display | 0.01 °C in the range -9.99…99.99 °C, else 0.1 °C | 0.01 °F in the range -9.99…99.99 °F, else 0.1 °F |
Reference temperature sensor | internal, fixed installation | |
Interface | RS485 (calibrator) to USB (PC) | |
Connectivity | MODBUS | |
Dimensions | ||
g Width
g Height g Depth |
210 mm
380 + 50 mm 300 mm |
|
Weight | Approx. 12.5 kg | |
Power supply | 100…240 VAC, 50 / 60 Hz | |
Power consumption | Approx. 375 W | |
Display | ||
Display | 2-line, 4-digit digital display, red / green, unit °C / °F |
You can request the necessary certificates, documentation, instructions and user guide for SIKA TPM165S using the special form.
You can buy SIKA TPM165S in UAE at a favourable price in «Texair FZCO» by leaving a request on the website, by phone +441225535040 or by email ceo@texair.eu. SIKA TPM165S price are subject to change depending on exchange rates. We deliver instruments by transport companies in United Arab Emirates, Kuwait, Qatar, Saudi Arabia, Saudi Arabia, Bahrain and other Middle Eastern countries.