+971 56 389 7151

Mon-Fri: 8:00 - 18:00 UAE

Email:

Michell QMA601LR Low Range Process Moisture Analyzer

Fast, High-Precision Moisture Measurement

  • Reliable low-range measurement from 0.02 to 100 ppmV
  • Accuracy of ±0.03ppmV or 10 % of reading
  • Maintenance free for 3 years
  • Built-in verification of customer process gas
  • IECEx, ATEX, UKCA, TR CU Ex certified for Exd flameproof, cQPSus certified for explosion proof
  • Intuitive color HMI with touchscreen keypad; no hot works permit required
  • 14 internal alarms
  • 21 predefined carrier gases and 3 user-defined gases
  • Proven Michell Instruments quality: 40 years’ expertise in moisture measurement built into the design
			int(21062)
NULL
		

Made in United Kingdom. brand Michell, model QMA601LR

International delivery to United Arab Emirates, Kuwait, Qatar, Saudi Arabia, Saudi Arabia, Bahrain, Oman and other Middle Eastern countries.
Our warehouses are located in UAE. Delivery to other countries (Europe, North and South America, Asia, Africa) is discussed individually.

price По запросу

Available

Available

Advantages

Only the best I&C brands

Our company is a UAE supplier of control and measuring instruments from well-known world brands

Advantages

Discounts for organisations

We sell high-tech equipment at best prices. We provide really advantageous conditions for partners

Advantages

Fast delivery across the Middle East

Supply to United Arab Emirates, Kuwait, Qatar, Saudi Arabia, Bahrain, Oman and other countries

Advantages

Professional advice on device selection

Our specialists are ready to advise you on any technical question. We can find an analogue of any device

Description of Michell QMA601LR

The QMA601-LR Moisture Analyzer is designed to provide reliable, fast and accurate measurement of ultra-low trace moisture content in natural gas in NGL fractionation and LNG liquefaction applications. It has a lower detection limit of 0.02 ppmV with reliable accuracy at sub-ppm trace moisture levels. The low-range variant is the result of Michell Instruments’ ongoing efforts to improve Quartz Crystal Microbalance technology for ultra-low trace moisture measurements down to 20ppbV. While measurements of quantities of moisture at ppbV levels stretch other technologies, the new QMA601-LR can offer reliability, simplicity, and greatly reduced cost of ownership from trusted and proven quartz crystal technology.

The analyzer requires only minimal routine maintenance to ensure low cost of ownership. It features an intuitive ‘through the glass’ touchscreen interface which is both easy to use and means that the analyzer can be interrogated and configured without the need for a hot works permit.

Applications

  • Molecular sieve dehydration of natural gas in:
    – Natural Gas in NGL fractionation
    – LNG liquefaction plants

Technical specifications of Michell QMA601LR

Performance Specifications
Measuring Technology Fast-Response Quartz Crystal Microbalance
Calibrated Range 0.02…100 ppmV, certified traceable to national humidity standards at NPL (UK) and NIST (USA)
Measurement Range 0.02…100 ppmV, trending to 2000 ppmV
Accuracy ±0.03 ppmV or ±10% of reading, whichever is greater
Repeatability ±5% of reading from 0.3 to 100 ppmV
Limit of Detection 0.02 ppmV
Available Units ppmV, ppmW, mg/Nm³, vapor pressure (Pa), dew point (°C/°F), lb/MMscf
Response Speed Close to instantaneous response to sample gas moisture changes
Automatic Calibration Internal moisture generator source, nominally 0.5 ppmV, calibrated traceable to NPL and NIST
Sensitivity 0.01 ppmV or 1% of reading, whichever is greater
Electrical Specifications
Supply Voltage 85…264 V AC, 47/63 Hz or 24 V DC (ATEX / UKCA / IECEx)
24 V DC only (cQPSus)
Alarms 1 x system alarm, volt-free change-over (FORM C)
3 x process alarms, selectable for various parameters, volt-free changeover (FORM C)
Analog Signals 2 x 4…20 mA or 1…5 V (selectable)
Maximum load resistance 500 Ω for 4…20 mA and minimum load of 1 MΩ for 1…5 V
Digital Communications RS485 Modbus RTU, Modbus TCP
Data Logging Available on analyzer (limited number of values) or via Application Software
Local Interface 7” color LCD with intuitive HMI
Electrical Connections M20 entries for cable glands
Expand Table

Operating Conditions

Inlet Pressure 2 barg
Outlet Pressure 1 barg
Sample Flow 300 ml/min total flow
Sample Gas Temperature 0…+100 °C
Operating Environment Analyzer only: +5…+45 °C up to 90 %rh
Analyzer in sampling system: -20…+55 °C up to 95 %rh (fitted with heater/thermostat and/or enclosure cooling as appropriate to maintain +5…+45 °C internal temperature)
Mechanical Specifications
Type GUB Flameproof Exd
Enclosure Lid & body: Cast copper-free aluminum LM25 (EN AC-42000), less than 0.6 magnesium
Glass window: Heat resistant, explosion proof, polyester coated, IP66, NEMA 4
Analyzer Gas Connections 1/8” NPT
Weight 35 kg without sampling system
Sample System Enclosure 316L stainless steel
Certifications
Hazardous Area Certifications ATEX/UKCA: II 2 GD Ex db IIB+H2 T6 Gb, Tamb -40 °C…+60 °C
IECEx: Ex db IIB+H2 T6 Gb, Tamb -40 °C…+60 °C
cQPSus: CLS I, Div 1, Group BCD T6 Tamb -25 °C…+55 °C
CLS I, ZONE 1, AEx db IIB + H2 T6 Gb Ex db IIB + H2 T6 Gb Tamb -20 °C…+55 °C
1Ex d IIB+H2 T4 Gb X, 1Ex tb IIIC 130 °C Db X and 1Ex d IIB+H2 T3 Gb X, 1Ex tb IIIC 195 °C Db X
TR CU: Certified for use in Russia
Expand Table

User Guide and Instruction manual for Michell QMA601LR

You can request the necessary certificates, documentation, instructions and user guide for Michell QMA601LR using the special form.

You can buy Michell QMA601LR in UAE at a price starting from По запросу $ in «Texair FZCO» by leaving a request on the website, by phone +441225535040 or by email . Michell QMA601LR price are subject to change depending on exchange rates. We deliver instruments by transport companies in United Arab Emirates, Kuwait, Qatar, Saudi Arabia, Saudi Arabia, Bahrain and other Middle Eastern countries.

Generate an article by selecting the parameters below
Article:
Quantity:
-
+
Clear
I know the article / Need help
Get Offer
The article is not fully generated. Please double-check the selected values.